Choose the measurement before choosing the instrument. An isothermal microcalorimeter holds a battery or cell at a controlled temperature and records heat flow, making it suitable for self-discharge, heat generated during cycling, parasitic reactions and thermal-management work. The correct configuration depends on your cell geometry and scale, minimum signal, maximum sustained heat, temperature program, electrical connections, throughput and safety enclosure. If the objective is runaway or abuse behavior, evaluate an ARC or other adiabatic system as a separate method rather than treating every calorimeter as interchangeable.
Start with the question your experiment must answer
Isothermal calorimetry is a measurement condition, not a battery form factor. The thermostat keeps the specimen at a defined temperature while the instrument measures thermal energy moving between the specimen and its surroundings. That lets you relate heat flow to an electrical or chemical process when the cell fixture and data system support simultaneous measurement. TA Instruments describes applications including self-discharge, charge and discharge, electrochemical calorimetry, parasitic reactions, structural events and safety assessment in battery research and quality work (TA Instruments overview).
Baseline self-discharge
For self-discharge, the useful signal may be small and persistent. Specify the lowest heat flow you must resolve, the duration of the run and the required baseline stability before considering a nominal sensitivity figure. A cell that is electrically open-circuit is a different protocol from a cell connected to a monitoring circuit, so include the intended wiring and any reference vessel in the measurement plan.
Heat during charge and discharge
Cycling studies need a calorimeter that can accommodate the battery, current-carrying leads and the electrical protocol without compromising the thermal measurement. Define charge and discharge currents, voltage limits, rest periods, cycle count and whether the calorimeter must exchange data with a cycler or potentiostat. The instrument’s heat range must cover both the low-heat rest periods and the highest sustained load in the cycle.
#1 Best Overall
- Powerful Turbo Fan:CAGIWIRU Electric Air Duster have reached speed up to 130000RPM,Can efficient remove dust and debris.With LED light and Three Adjustable Speed to meet with different cleaning tasks.
- With Attached USB C Fasting Charging cable.Charing 3 hours,Enjoy up to 240 minutes of use on the lowest setting, For third speed can use for 25mins.with four charging options to meet with your needs.
- Wide Application: attached 5 different nozzles,making it suitable for a variety of scenes, such as car,pet, pc, keyboards, and other electronic devices. It also serves for office and home clean duster.
- Handy Design & Portable:Cordless air duster electric nozzle is Light weight and small size, design and comfortable to hold and space saving, convenient to carry around. You can put it in your room or inside the car without taking up space.
- Economical &Reusable: Built durable materials with a long-lasting battery,our CAGIWIRU air duster it's a sustainable alternative to disposable air cans.
Parasitic and irreversible reactions
Formation, aging and storage experiments often seek heat that is not explained by the intended reversible electrochemistry. These tests place unusual demands on baseline drift, reference matching, temperature equilibration and run duration. Ask the supplier how calibration, blank subtraction and reference vessels are handled for the exact cell fixture you will use; a brochure sensitivity value does not establish the quality of a long, low-signal baseline.
Thermal properties and management
If you need heat capacity, thermal resistance or a heat-balance input for a thermal model, the specimen geometry and mounting hardware become part of the measurement. A fixture that is appropriate for a coin cell may not represent the thermal path of a pouch cell, module or pack. Document the mechanical interfaces and any compression, insulation or heat-spreader parts that will remain in the test.
Safety or abuse response
First determine whether the experiment is controlled heat generation or an uncontrolled exothermic event. H.E.L. distinguishes isothermal calorimetry for heat generation and efficiency from ARC-type adiabatic calorimetry for thermal-runaway and safety testing (H.E.L. battery-testing technologies). NETZSCH likewise describes isothermal cycling for heat-flow and life characterization and ARC for runaway scenarios (NETZSCH battery applications). If pressure rise, venting, abuse or self-heating under near-adiabatic conditions is the main question, make ARC capability, containment and hazard review a primary procurement requirement.
The Tool Desk
Outbyte PC Repair FREERepair Windows errors before they cause bigger problemsFix Now →Outbyte Driver Updater FREEScan for outdated or missing drivers - takes under a minuteDriver Scan →Match the chamber to the cell, module or pack
List every specimen you expect to test during the instrument’s service life, not just the first cell chemistry. Confirm the usable chamber volume, insertion path, lifter or ampoule design, electrical clearance and space for a reference vessel. TA describes removable sample cells from below 1 mL to 125 mL depending on configuration and lists battery lifters for common coin, pouch and 18650 formats (TA isothermal microcalorimeters). NREL’s battery calorimeter documentation separates cell, module and large-volume IBC configurations (NREL Isothermal Battery Calorimeters).
| Specimen scale | Questions to resolve before buying | Why it changes the instrument choice |
|---|---|---|
| Coin or small laboratory cell | Does the holder position the cell reproducibly? Are leads and a reference vessel included? What is the lowest heat signal of interest? | Small thermal signals make baseline noise, drift and fixture heat capacity critical. |
| Cylindrical cell, including 18650-size formats | Does the lifter accept the diameter and length? Can the leads exit without touching the chamber? Is the cell restrained against movement? | The insertion fixture and electrical clearance can limit usable cell dimensions even when chamber volume appears adequate. |
| Pouch cell | What pouch footprint, thickness and compression hardware are supported? Can the lid close without bending tabs? | Mechanical contact and tab routing affect both thermal coupling and electrical reliability. |
| Module | What is the maximum mass, volume and sustained heat? Are multiple electrical channels available? | A module may exceed the heat capacity and heat-load limits of a high-sensitivity small-cell setup. |
| Pack or large assembly | Is there a large-volume chamber, safe insertion procedure, ventilation or containment plan? | Large specimens require a different thermal mass, safety envelope and facility workflow. |
Do not infer pack capability from a cell holder. The chamber, thermal control loop, wiring and safety enclosure all scale with the specimen. Obtain a drawing showing internal dimensions and cable routing for the exact configuration in the quotation.
Specify both sensitivity and maximum heat
Two limits matter: the smallest heat flow that must be distinguished from baseline noise and the largest sustained heat that must be measured without saturation or unsafe temperature rise. They are not interchangeable. A system optimized for micro-watt signals may not be suitable for a module producing tens or hundreds of watts.
Rank #2
- Screen Cleaner is Your Screen’s New Bestfriend! - EVEO is proud to present a screen cleaner that’s perfectly suited for your TV. Meet your new screen cleaner, computer screen cleaner, laptop cleaner, iPad cleaner, Macbook,tv cleaner or any other electronic device. Our screen cleaner spray and wipe along with the included plush microfiber cloth, easily removes dust, fingerprints & other smudges on the screen’s surface. Clean effectively for a streak-free screen
- Gentle on your Screens - Our screen cleaner spray and wipes is compatible to be used even for the most sensitive LCD, LED, CRT, and OLED screens in proper use. Be at ease knowing that this product is compatible with all the major brands: LG, SONY, Samsung, Sharp, iPad, iPhone, Android screens, Kindle, PC monitors - we’ve got you covered! Can be used as LCD screen cleaner, tv cleaner, smart tv screen cleaner, and more
- Premium Super Soft Plush Microfiber Cloth - Included in this efficient screen cleaning kit. This screen cleaner spray and wipe has our signature plush microfiber cloth specially designed to comply with screens and monitors and leave them streak-free. Has been shown to effectively absorbs stubborn stains, zero streaks, and no fiber shedding. Always use this together with our screen cleaner for the best results
- Easy-to-Use Screen Cleaner Kit + microfiber cleaning cloth - The EVEO Screen Cleaning Kit will easily be your new favorite tool when it comes to taking care of the things you love most, quick and easy solution for a streak-free screen. Simply screw on the sprayer gun on the screen cleaner bottle. Spray the cleaning liquid onto the microfiber cloth, then wipe the screen until desired cleanliness & shine is achieved.
- Quality Screen Care Kit Worth Purchasing - Enjoy a convenient Screen Cleaner anytime, anywhere you need it. Your satisfaction is our top priority. We stand behind the quality of our products and that’s why, you shouldn’t be worrying at all. We manufacture the Best Screen Cleaner. and can be used for laptop cleaning kit, laptop screen cleaner, tv screen cleaner for smart tv. MacBook screen cleaner, monitor cleaner or screen cleaner spray for electronic devices with microfiber cleaning cloth
| Configuration or example | Published heat figure | How to interpret it |
|---|---|---|
| NREL IBC 284 cell configuration | 50 W maximum constant heat generation (NREL, 2015) | An example specification for the cell-scale IBC configuration, not a current market-wide limit. |
| NREL Module IBC | 150 W maximum constant heat generation (NREL, 2015) | Shows why module testing needs a higher heat-load class than a small-cell setup. |
| NREL Large-Volume IBC | 4,000 W maximum constant heat generation (NREL, 2015) | An example large-volume specification; verify present availability and safety provisions. |
| NETZSCH MMC274 Nexus ARC module example listed by NIMS | 10 μW heat-flux sensitivity (NIMS Battery Research Platform; year not stated on the equipment page) | A sensitivity figure for that facility configuration, not a cross-vendor ranking or a guaranteed specification for every MMC274 installation. |
The NREL values come from a FY2015 specification sheet, so use them as an engineering reference and ask whether the configuration is still offered. The NIMS page describes a facility example rather than a universal commercial package (NIMS ARC equipment page). For your request for quotation, provide a heat-flow envelope: minimum signal, typical signal, maximum continuous signal, short transients and expected run duration.
What’s actually slowing this PC down?
Pick the symptom - the matching free tool is one click away.
Ask for noise and baseline data, not only resolution
“Sensitivity” can describe a detection threshold under a particular setup; it does not tell you how stable the baseline will be over a 24-hour self-discharge run. Request blank and reference-cell data acquired with the proposed vessel, lead arrangement and temperature setpoint. Ask how the supplier defines noise, drift, response time and usable dynamic range, and whether the quoted values apply with electrical feedthroughs installed.
Check temperature range, stability and equilibration
Compare the instrument’s setpoint range with your real protocol, including storage, cycling and abnormal-condition temperatures. NREL lists −30 to 60 °C for its cell and module configurations and −40 to 100 °C for its large-volume configuration (NREL specification sheet). TA describes temperature control within 100 μ°C over a minimum of 24 hours for its TAM platform; confirm that the statement applies to the model, vessel and accessory combination in your quotation (TA platform information).
Temperature range alone is insufficient. Ask for:
- Setpoint accuracy and short-term stability at each required temperature.
- Long-term drift over the duration of your longest experiment.
- Typical equilibration time after inserting a cell, changing temperature or connecting electrical leads.
- Calibration method and the interval between calibrations.
- Whether the stated stability is measured at an empty chamber, with a reference vessel or with a battery under load.
Plan the thermal sequence around equilibration. A high-current cycling test started before the cell and reference have reached the same thermal state can obscure the first part of the heat trace. Your standard operating procedure should define the stabilization criterion and the point at which data collection begins.
Make electrical integration a core specification
For in-operando calorimetry, electrical integration is part of the measurement system, not an optional convenience. Specify the number of channels, current and voltage limits, connector type, isolation, cable resistance, feedthrough sealing and synchronization between electrical and heat-flow data. Confirm that the calorimeter can work with your existing cycler or potentiostat and that the software exports timestamps in a format your analysis tools can use.
Outdated Drivers Are Slowing You Down
One free scan finds every outdated or missing driver and matches the right update for your exact hardware.Free scan · exact hardware matchPC Slower Than It Used to Be?
A free scan shows the junk files, broken settings and background clutter dragging Windows down - then fixes them in one click.Free scan · Windows 10 & 11TA’s named integrated example combines a TAM IV with a BioLogic VSP-300 potentiostat, and its general TAM information lists electric-charge capability among available accessories (TA battery application note; TA product information). Treat that combination as an example of integration, not evidence that every configuration supports your current, voltage or control protocol.
Rank #3
- 【Large size】:Measuring 800 x 300 mm/ 31. 5×12 inches, which is wider and taller than others. Large mouse pad can perfectly fit your keyboard and mouse of any size, and there is also ample space for peripherals such as phones, tablets, etc.
- 【LED backlight mouse pad】:Blade Hawks RGB mouse pad provides 7 static modes and 7 dynamic modes will give you the options you are looking for. Power-off memory function remembers the last lighting mode you selected. The super glow fiber Chroma will give you a colorful gaming setup that you want.
- 【Anti-slip rubber base】:Made from Natural rubber, this gaming mousepad will firmly grip your desktop, allowing you to enjoy the ultimate precision in the games you are most passionate about. It also comes with a texture that enhances this ability even further.
- 【Smooth and waterproof surface】: Micro-textured cloth surface, for extremely precise mouse control and a smooth movement. When liquid splashes on the gaming mouse pad, it forms water droplets and slides down. Will not delay your work or gameplay.
- 【Plug and play】: This mouse and keyboard pad is powered by USB, plug and play, no driver required. One button control light modes. Press one time to change the lighting mode, press twice to change the brightness, press and hold about 3 seconds turn ON/OFF.
Questions for the electrical interface
- What continuous and peak current can the feedthrough and fixture carry?
- What voltage range and insulation rating apply at the intended temperature?
- Can the instrument record heat and electrical channels on one clock, or must you synchronize files afterward?
- Does the control software support charge, discharge, rest and cycling commands, or does an external cycler control the sequence?
- How are failed-cell disconnection, over-temperature shutdown and emergency stop handled?
- Can you replace cables and contacts without recalibrating the calorimeter?
Separate isothermal calorimetry from ARC and adiabatic testing
In an isothermal test, the instrument deliberately holds the specimen near a chosen temperature so heat generation can be compared across operating conditions. In an ARC or adiabatic test, the system seeks to follow the specimen’s self-heating and characterize the transition toward thermal runaway. Those methods answer different questions and impose different containment and hazard requirements.
| Decision point | Isothermal calorimetry | ARC or adiabatic calorimetry |
|---|---|---|
| Primary question | How much heat does the battery produce during a controlled protocol? | When does self-heating begin, and how does an exothermic event progress? |
| Typical work | Self-discharge, cycling heat, parasitic reactions, efficiency and thermal-model inputs. | Runaway, abuse, venting and safety response. |
| Thermal condition | Controlled, near-constant specimen temperature. | Adiabatic or near-adiabatic tracking of self-heating. |
| Procurement emphasis | Baseline performance, heat range, temperature stability, electrical integration and throughput. | Detection logic, pressure and gas handling, containment, emergency response and hazard engineering. |
Some platforms or facilities combine modes, but verify which module, vessel and test conditions are included. A multi-mode label does not guarantee that a single fixture can safely perform both a low-signal cycling experiment and a high-energy abuse test.
Plan throughput, atmosphere and daily workflow
Throughput is more than the number of holes in a sample rack. Count the time to assemble cells, install lifters, equilibrate temperature, run a calibration, remove specimens and clean or inspect the chamber. TA lists adjustable throughput up to 48 samples depending on configuration and accessory options, including humidity, perfusion, pressure and electric-charge accessories (TA product information). Confirm whether the quoted number means simultaneous measurements, sequential positions or a particular vessel set.
Environment and accessories
- Atmosphere: Specify inert gas, controlled humidity or ambient operation and identify where gas enters and exits.
- Pressure: If pressure measurement or controlled pressure is required, confirm rated vessels, sensors and relief paths.
- Liquids and perfusion: Verify compatible materials, tubing dead volume and whether flow heat is included in calibration.
- Reference vessels: Match reference volume, mass and wiring to the sample side so the differential signal remains meaningful.
- Data: Confirm raw-data access, metadata fields, audit trails, export formats and user permissions.
Facility and service requirements
Before ordering, identify bench space, electrical service, ventilation, gas supplies, network access and emergency procedures. Ask who installs the system, who performs acceptance testing, how often calibration is required, what preventive maintenance includes and how quickly service parts are available in your region. Request training for cell assembly, safe insertion, calibration checks, method development and failure response.
Use a weighted comparison rather than a brand-first shortlist
Create a requirements sheet and score each configuration against the same evidence. The following axes keep a quotation focused on your experiment:
| Axis | Evidence to request | Suggested pass condition |
|---|---|---|
| Method | Isothermal, ARC, adiabatic or combined; supported vessels and modes. | The primary method directly answers the stated research question. |
| Specimen | Drawings, mass and volume limits, supported coin, cylindrical, pouch, module or pack fixtures. | Every planned specimen fits with required wiring and safety clearance. |
| Heat range | Minimum detectable signal, noise, drift, maximum continuous heat and transient handling. | The full protocol envelope fits without saturation and with documented baseline performance. |
| Temperature | Range, stability, accuracy, equilibration time and calibration conditions. | All setpoints and stabilization criteria are met with the intended fixture. |
| Electrical integration | Current, voltage, channels, feedthroughs, cycler compatibility and timestamp synchronization. | The complete charge/discharge sequence runs without manual data reconstruction. |
| Throughput | Simultaneous positions, changeover time, accessories and operator steps. | Capacity matches the number of samples and turnaround your program requires. |
| Safety and service | Containment, shutdowns, installation, training, calibration and response time. | The facility can operate the system under an approved risk assessment. |
| Cost of ownership | Configuration-specific quote, consumables, calibration, service and infrastructure. | Budget includes the complete installed system, not only the analyzer. |
The published sources do not provide comparable market-wide prices or total-cost figures. Request itemized, configuration-specific quotes rather than using an old list price or estimating from a neighboring model.
Rank #4
- 【Softer and More Comfortable】Vaydeer wrist rest has unique diamond pattern, which is the combination of softness and aesthetics. The materials of wrist rest are improved into higher quality memory foam and covered with silky smooth lycra. The computer wrist rest makes you as comfortable and cushiony as like rest your wrists on clouds.
- 【Ergonomic Wrist Saver】The wrist rests for keyboard and mouse comes with a 17.32×3.15×0.83 inch keyboard wrist pad and a 5.94×3.15×0.83 inch mouse wrist support. Based on ergonomic design, the unique concave shape is the perfect fit for your wrist joints. The wrist rest pad fits most computer keyboards and laptops, improve hand and wrist posture, release your wrist and arm stress.
- 【Non-Slip Rubber Bottom】Featuring an anti-skid silicone base on the bottom, this wrist keyboard support stays firmly in place on your desk, preventing the padding from sliding around, ensuring stable and consistent wrist support during extended computer sessions.
- 【Better Experience & Pain Relief】Our keyboard arm rest is beneficial to alleviate the soreness caused by direct contact and friction between your arm and a hard desk surface, reducing the risk of wrist fatigue or carpal tunnel. The soft texture of memory foam can evenly distribute the pressure around your wrists and provide good support with just enough give.
- 【Helpful in Multiple Scenarios】Whether you're working, studying, writing, typing, gaming, this keyboard and mouse rest combo is an essential accessory to add comfort and support to your hands and wrists. It’s also a great gift for men, women, family, friend, coworker, gamer, teacher, etc.
A practical procurement sequence
- Write the measurement brief. State the chemistry, cell format, electrical protocol, temperature setpoints, run duration, expected minimum and maximum heat, atmosphere and safety hazards.
- Separate routine and exceptional tests. Identify which work is low-signal isothermal measurement and which work involves abuse, pressure or runaway. Decide whether one platform genuinely needs to cover both.
- Draw the specimen and wiring. Include dimensions, tabs, compression, lifters, reference vessel, cable bend radius and any sensor that must remain inside the chamber.
- Define acceptance tests. Include blank baseline, reference-cell response, temperature stability, electrical synchronization, maximum-load demonstration and emergency shutdown checks.
- Send the same specification to suppliers or facilities. Require each response to identify the exact model, vessel, fixture, software, accessories, exclusions and current availability.
- Review the safety case. Have laboratory safety personnel approve containment, gas handling, pressure relief, fire response and operator training before placing an order.
- Compare installed ownership cost. Add calibration, service, consumables, replacement fixtures, facility modifications, training and downtime to the instrument quotation.
- Run a pilot method. Use representative cells to verify equilibration, baseline drift, electrical synchronization and heat-range margin before committing to a large campaign.
Common selection mistakes and how to avoid them
Choosing by the smallest published sensitivity
A single sensitivity number does not describe noise, drift, dynamic range or performance with electrical leads. Require data from the proposed configuration and compare the complete heat-flow envelope.
Recommended Free Tools
Assuming a cell instrument can test a module
Cell, module and large-volume systems can have very different chamber capacities and heat limits. Verify dimensions, mass, maximum continuous heat and safety containment for the largest specimen you will actually run.
Adding electrical integration after purchase
Feedthroughs, insulation, current limits and synchronization affect the calorimeter design. Specify them before the quotation and include them in acceptance testing.
Confusing temperature range with usable protocol
A nominal range does not tell you how long equilibration takes or how stable the temperature remains under load. Ask for stability and equilibration evidence at your setpoints with the intended vessel.
Using isothermal equipment for a runaway question
If the event of interest is self-heating, pressure rise or abuse, evaluate ARC or adiabatic capability and the associated containment instead of relying on an isothermal heat-generation setup.
Quick wins for a faster PC:
Clear out junk files and repair common Windows errorsFree Scan →Fix the driver behind crashes, sound loss and screen glitchesFind Drivers →Repair Windows errors before they cause bigger problemsFix Now →Budgeting only the analyzer
Fixtures, lifters, reference vessels, electrical hardware, gas or humidity accessories, calibration, installation and training can determine whether the laboratory can use the system safely and productively. Demand an itemized installed quote.
Best Value
- 【No Drill Mounted】Cable management tray can be assembled in as little as 3 minutes or less-simply. Side clips easily clampe the edge of desk, pass through the holes on the sides, and secure excess cables with baskets. Desk cable management supports inward or outward installation to meet your needs in different scenarios, and provides you with great convenience when collecting and organizing wires.
- 【Rubber Pad Protects Desk】Wire organizer under desk built-in soft rubber pads on the clamp protect your desk from scratches, so you will not damage your furniture or office.
- 【Sturdy and Beautiful】Cord management under desk is made of high-quality carbon steel and holds about 15 lbs. The wire basket has 2 holes on each side of the tray to enable your cables to pass freely. Cable baskets are suitable for desk thickness use from 0.4" to 2.4".
- 【Space Saving & Keep Organized】Desk cable management tray holds all your power cords, outlet strips, USB hubs and computer transformers hidden from your feet, and the data cables will not enter the vacuum cleaner to keep your desk clean and tidy. Desk wire organizer is also suitable for use in the kitchen and outdoors.
- 【Reduce Safety Risks】Cable tray under desk keeps all plugs away from your kids, no more worry about tripping. Using 2 holes wire tray saves extra space, helps you easily cabling and wire protection. The mesh design is not easy to accumulate dust, but avoids safety accidents caused by messy wires.
Example decisions
Low-heat self-discharge of coin cells
Prioritize baseline noise, long-term drift, stable temperature, repeatable small-cell fixtures and a reference vessel. Electrical integration may be limited to open-circuit monitoring, but cable routing and feedthrough leakage still need to be specified. A large-volume heat capacity is unnecessary unless future work justifies it.
Heat generation during pouch-cell cycling
Prioritize pouch dimensions, tab routing, compression hardware, current and voltage limits, synchronized cycler data and a heat range that includes both rest and peak cycle loads. Ask for an equilibration procedure because pouch mass and fixture contact can lengthen stabilization.
Module-level thermal characterization
Start with the module’s maximum sustained and transient heat, mass, dimensions and cable count. NREL’s example limits—50 W for its cell IBC, 150 W for its Module IBC and 4,000 W for its Large-Volume IBC—illustrate why the chamber class must be selected from the specimen scale rather than from sensitivity alone (NREL documentation).
Do these 3 things before closing this tab:
1Clear out junk files and repair common Windows errors2Fix the driver behind crashes, sound loss and screen glitches3Repair Windows errors before they cause bigger problemsRunaway or abuse investigation
Define the abuse trigger, pressure and gas hazards, containment, emergency shutdown and post-test handling first. Then evaluate ARC or adiabatic systems and confirm that the proposed vessel and detection logic are appropriate for the cell energy and event severity.
What to request in the final quotation
- Exact analyzer model and configuration identifier.
- Chamber drawings and supported specimen geometries.
- Heat-flow range, noise, drift, response and maximum continuous heat under stated conditions.
- Temperature range, stability, accuracy and equilibration specifications for the proposed vessel.
- All electrical feedthrough, cycler or potentiostat hardware and synchronization method.
- Number of simultaneous samples and the definition of throughput.
- Atmosphere, humidity, pressure, perfusion and reference-vessel accessories.
- Calibration, installation, training, acceptance testing and preventive-maintenance scope.
- Safety enclosure, shutdown functions, ventilation or gas requirements and facility responsibilities.
- Itemized purchase, consumable, service and replacement-fixture costs, with quotation date and validity.
The best choice is the configuration that covers your complete measurement envelope with documented baseline behavior, safe specimen handling and a workflow your laboratory can repeat. Start from the experiment, then make the chamber, heat range, temperature control, electrical interface and safety system prove that they fit it.