Microcrystal electron diffraction (MicroED) can determine structures from crystals far too small for conventional X-ray diffraction, including some experiments that use only a tiny fraction of a milligram of starting material. It is not, however, a mass-only solution: the sample must form ordered three-dimensional crystals with suitable thickness, the crystals must be prepared without damaging them, and a cryo-transmission electron microscope (cryo-TEM), detector and experienced operators must be available.

The practical question is therefore not “How many milligrams are required?” but “Can this material produce diffraction-quality crystals, and can they be mounted and measured with the available MicroED workflow?”

What MicroED determines—and why it helps with tiny samples

MicroED is an electron-diffraction method for solving crystal structures. A beam of electrons passes through very small three-dimensional crystals inside a transmission electron microscope. As the crystal is rotated, the resulting diffraction patterns are recorded and combined to calculate a structure. The method has been applied to proteins, peptides and small molecules.

Special offer. See more information about Outbyte and uninstall instructions. Please review EULA and Privacy policy.

Its main advantage is crystal size. Conventional single-crystal X-ray diffraction generally needs a crystal large enough to mount and illuminate effectively. MicroED can use microcrystals that are too small for that approach. Thermo Fisher Scientific states that crystals only 100 nm in size can readily be analyzed in an illustrative MicroED context. That figure describes a demonstrated capability, not a universal lower limit for every compound, instrument or preparation.

#1 Best Overall
Elebase USB to USB C Adapter for iPhone 18 Pro Max,USBC Car Charger Adapter | 4Pack Type C Female to A Male Cable Converter for Apple Duo 17 16 15 14 Plus,iWatch Watch 12 11 10,iPad 9 8 Air,Airpods 5 4
  • Read Before You Buy — No Video Output: These adapters support charging and USB 2.0 data transfer, but cannot transmit video signals. Except for standard USB webcams (which use USB data only), they are not compatible with HDMI/DisplayPort cables, video-capable USB-C hubs, or docking stations with video output.
  • Convert USB-A Ports to USB-C: Designed to connect USB-C earphones, cables, flash drives, card readers, and other USB-C accessories to standard USB-A ports. Plug-and-play with no drivers or software required.
  • Aluminum Alloy Housing: Built with a sturdy aluminum alloy shell that aids in heat dissipation and protects against daily wear and scratches. Designed to maintain a stable and secure connection.
  • Compact & Travel-Friendly: The ultra-compact design allows the adapter to stay plugged into your device without blocking adjacent ports or adding bulk, reducing wear and tear on your original USB ports.
  • 12-Month Warranty: Backed by a 12-month manufacturer warranty for peace of mind. Designed to meet strict quality control standards for reliable everyday performance.

Electrons interact strongly with matter. That strength makes very small crystals detectable, but it also means thickness is a central constraint: an overly thick crystal can produce complicated or unusable scattering. A well-ordered crystal that is thin enough for the beam is more valuable than a larger but poorly ordered or excessively thick particle.

MicroED does not remove the need for crystallization. You still need crystals, and their internal order must be good enough to produce interpretable diffraction. If a target remains amorphous, forms only disordered precipitate, or produces crystals with unsuitable geometry, reducing the sample mass will not make MicroED succeed.

There is no universal minimum mass

Sub-milligram work is possible, but “sub-milligram” is not a specification that guarantees a structure. The amount of starting material, the number of crystals produced, crystal volume, diffraction quality and losses during preparation all affect feasibility.

Special offer. See more information about Outbyte and uninstall instructions. Please review EULA and Privacy policy.

Thermo Fisher Scientific describes a particular MicroED structure obtained with approximately 10-12 grams of sample. This is an example from a vendor overview, not a general minimum-mass threshold or a promise that a new target can be solved from that amount. The mass may refer to material represented in a specific experiment rather than a transferable recipe for every sample type.

For planning, separate three quantities:

  • Starting material: the protein, peptide or compound you can make or purify before crystallization.
  • Crystallized material: the fraction incorporated into ordered crystals rather than remaining in solution or precipitating poorly.
  • Material actually measured: crystals that can be located, are thin and isolated enough for the beam, and produce data over a useful angular range.

A small starting mass can be sufficient when crystallization is efficient and many suitable microcrystals are available. Conversely, a larger preparation may fail if crystals are too thick, damaged during handling or poorly ordered. Report any mass as an experiment-specific amount with its preparation and measurement conditions.

Rank #2
Anker USB-C Hub, 5-in-1 USB Hub for Laptops, 4K HDMI Multiport Adapter | USB C Dongle with 90W Max Power Delivery, 3×5Gbps USB A Data Ports, for MacBook, Dell, and More (Charger Not Included)
  • 5-in-1 USB-C Hub: Experience comprehensive connectivity featuring a Power Delivery input, two USB-A 2.0 ports, a USB-A 3.0 port, and an HDMI port. (Note: The USB-C power delivery input port is only for connecting an external wall charger to power your laptop and cannot power peripheral devices.)
  • 90W Pass-Through Charging: Achieve optimal charging with 90W pass-through power to your laptop, supported by a total input of 100W, with the hub reserving 10W for operational efficiency. (Note: Wall charger not included.)
  • Quick Data Transfers: Accelerate your productivity with rapid data transfers using a high-speed 5Gbps USB 3.0 port and two 480Mbps USB 2.0 ports.
  • 4K HDMI Display: Enhance your visual experience with a hub capable of delivering 4K resolution at 30Hz in both mirror and extend modes. Please note that this hub is compatible with MacBook (macOS 12 and newer), Windows 10 and 11, ChromeOS, and laptops equipped with DP Alt Mode and Power Delivery. Note: This device is not compatible with Linux.
  • What You Get: Anker USB-C Hub (5-in-1, 4K HDMI), welcome guide, 18-month warranty, and our friendly customer service.

Check whether your sample fits MicroED

Before committing scarce material, ask the facility or crystallographer to assess the following factors.

The target forms crystals

MicroED is a crystal-diffraction method, not a replacement for crystallization. For proteins, the starting point is usually a suspension of microcrystals in solution. For small molecules, crystals may be obtained as a suspension or as a dry powder containing submicron crystals. If only amorphous material is available, a crystallization optimization step is needed first.

Special offer. See more information about Outbyte and uninstall instructions. Please review EULA and Privacy policy.

Crystal order is adequate

Particles that look crystalline under a microscope can still be too disordered to yield useful spots. Screening is required to identify crystals with consistent diffraction. A preparation containing many particles is not automatically better than one containing fewer, well-ordered crystals.

Thickness is appropriate

Because electrons scatter strongly, thickness can limit data quality. Very thick crystals may create multiple scattering or overlapping intensities that complicate processing. The acceptable thickness depends on composition, crystal habit, microscope conditions and the collection strategy; there is no single size cutoff that applies to all targets.

The sample can survive its required state

Macromolecular crystals must remain hydrated. The International Union of Crystallography’s 2025 “Standards for MicroED” warns that dehydration harms crystallinity and diffraction. Protein workflows therefore preserve water by vitrifying the sample. Some small-molecule crystals, by contrast, can be dried or examined at room temperature when their properties allow it.

Rank #3
Sale
Anker USB C Hub, 7in1 Multi-Port USB Adapter, 4K@60Hz USBC to HDMI Splitter | 85W Max PD, 2 USB 3.0 & 1 USBC Data Ports, SD/TF Card Reader, for Laptops/Mac and Other Type C Devices (Charger Not Included)
  • Sleek 7-in-1 USB-C Hub: Features an HDMI port, two USB-A 3.0 ports, and a USB-C data port, each providing 5Gbps transfer speeds. It also includes a USB-C PD input port for charging up to 100W and dual SD and TF card slots, all in a compact design.
  • Flawless 4K@60Hz Video with HDMI: Delivers exceptional clarity and smoothness with its 4K@60Hz HDMI port, making it ideal for high-definition presentations and entertainment. (Note: Only the HDMI port supports video projection; the USB-C port is for data transfer only.)
  • Double Up on Efficiency: The two USB-A 3.0 ports and a USB-C port support a fast 5Gbps data rate, significantly boosting your transfer speeds and improving productivity.
  • Fast and Reliable 85W Charging: Offers high-capacity, speedy charging for laptops up to 85W, so you spend less time tethered to an outlet and more time being productive.
  • What You Get: Anker USB-C Hub (7-in-1), welcome guide, 18-month warranty, and our friendly customer service.

There is a realistic data-collection plan

Useful diffraction requires more than one attractive image. The crystal must be accessible for rotation, separated enough from grid bars and neighboring crystals, and measured through a sufficiently complete angular range. Incomplete angular data can make indexing, integration and structure determination substantially harder.

Free tools Windows power users keep installed

One-click scans. No signup required.

Special offer. See more information about Outbyte and uninstall instructions. Please review EULA and Privacy policy.

Protein MicroED: a practical workflow

A published protocol, “MicroED sample preparation and data collection for protein crystals,” describes a cryogenic workflow. Exact concentrations, blotting conditions and microscope settings must be adapted to the sample and facility.

  1. Generate microcrystals in solution. Use a crystallization condition that produces ordered protein crystals rather than amorphous precipitate. Keep the suspension homogeneous enough to sample representative crystals.
  2. Choose compatible EM grids. The protocol uses holey-carbon or continuous-carbon coated electron-microscopy grids. Grid mesh, coating and support-film choice should be confirmed with the facility because they affect crystal retention, background and imaging.
  3. Apply the crystal suspension. Deposit a small amount of the suspension onto the prepared grid. The goal is to place crystals over usable support areas while avoiding heavy aggregation.
  4. Blot away excess liquid. Remove enough liquid to form a thin film around the crystals without drying the macromolecular sample. Excessive blotting can dehydrate or dislodge crystals; insufficient blotting can leave ice that is too thick for screening.
  5. Plunge-freeze for vitrification. Rapidly freeze the grid so the remaining water becomes a glass-like, non-crystalline state. Vitrification preserves hydration and limits movement during cryo-TEM examination.
  6. Screen in a cryo-TEM. Locate crystals, check their morphology and test diffraction. Crystals that are well defined and separated from grid bars or neighboring crystals are easier to screen and collect from.
  7. Select crystals with useful diffraction. Do not choose solely by apparent size. Favor particles that are thin enough, intact and sufficiently ordered for interpretable patterns.
  8. Collect continuous-rotation data. Rotate the selected crystal while recording diffraction with a high-speed detector. The accessible angular range and crystal stability determine how complete the data can be.
  9. Process and assess completeness. Index and integrate the frames, evaluate resolution and completeness, and determine whether the data support structure solution. Missing wedges or other incomplete angular coverage can become the limiting factor even when individual spots look strong.

Protein MicroED therefore requires more than a microscope booking. It combines grid preparation, controlled vitrification, cryogenic handling, crystal screening, diffraction collection and specialized data processing. A facility should review the sample state and expected crystal dimensions before you ship or consume irreplaceable protein.

Small-molecule samples follow a different preparation path

Small molecules can sometimes be handled more simply than proteins. The IUCr standards describe both liquid-suspended samples and dry powders containing submicron crystals. Depending on the compound, crystals may be dried or analyzed at room temperature rather than vitrified.

That flexibility does not eliminate the need for an appropriate crystal. Drying can alter a hydrate, polymorph or other solid form, so the facility must confirm that the intended preparation preserves the form you want to identify. A powder with a mixture of phases may produce overlapping or ambiguous diffraction, even if individual particles are crystalline.

Special offer. See more information about Outbyte and uninstall instructions. Please review EULA and Privacy policy.
Rank #4
Sale
UGREEN USB to USB C Adapter Combo 4-Pack, 10Gbps USB C Converter Space Gray | Data transfer & Fast charging for MacBook, iPhone 17/16/15, iPad, Samsung Galaxy, Apple Watch
  • Dual Converters, Infinite Potential:Includes 2× USB C male to USB A female adapters and 2× USB A male to USB C female adapters. Perfect for a wide range of uses—tablets with Bluetooth keyboards, expand USB ports on macbook, and more. Two different converters for all your daily needs
  • Next-Level 10Gbps & 3A Charging: No more slow 480Mbps, this usb to usb c adapter has a transfer speed of up to 10Gbps, allowing you to do more transferring in less time. This usb adapter fits both USB A and USB C charger, supporting up to 3A fast charging
  • Upgraded Exquisite Craftsmanship: With an aluminum alloy housing and metal connector, the usbc to usb adapter is extremely durable and sturdy. Rigorously tested to withstand more than 10,000 times of plugging and unplugging, ensuring long-lasting performance
  • Broad Compatible: The usb c to usb adapter widely supports all USB C/ USB A devices like laptops, tablets, cellphones, car chargers, and phone chargers. Such as compatible with MacBook Pro/Air 2023/2022, Thunderbolt 4/3 Devices,Apple MagSafe Watch 9/8/7/SE/Ultra, iPad Pro 2022/2021, Samsung Galaxy S23/S20/S10, and iPhone 17/16/15 Pro. Plug and play
  • Please Note: To reach 10Gbps speed, keep the cable under 3.3 ft. For USB A Male to USB C adapters, try flipping the USB C connector. USB C Male to USB A adapters support bidirectional 10Gbps transfer within 3.3 ft

When submitting a small-molecule sample, state whether it is a suspension, a dry powder, a known polymorph, a hydrate or a material with uncertain composition. The operator can then select a mounting and screening approach that minimizes transformation before measurement.

MicroED compared with other crystallographic routes

No single method is best for every sub-milligram project. Compare the options by crystal dimensions, sample handling, infrastructure and expected data completeness.

Decision factor MicroED Conventional single-crystal X-ray diffraction Serial X-ray diffraction
Crystal size Designed for very small three-dimensional crystals, including crystals too small for conventional X-ray mounting; dimensions remain sample- and instrument-dependent. Usually needs a larger single crystal that can be mounted and exposed effectively. Uses many crystals, with the required number depending on the delivery and collection setup.
Thickness constraint Critical because strong electron scattering can make thick crystals difficult to interpret. Crystal size and X-ray absorption are important, but thickness is managed differently from electron transmission. Each crystal still needs suitable diffraction quality; the aggregate experiment can tolerate variation across crystals.
Material quantity Published examples show extremely low amounts can suffice, but there is no universal mass threshold. Often requires enough material to grow a mountable crystal. May require many crystals even when each is small.
Sample state Proteins are generally hydrated and vitrified; some small molecules can be dry or room-temperature. Depends on the crystallography setup and cryoprotection strategy. Depends on the injector, fixed-target or other delivery method.
Infrastructure Cryo-TEM, suitable diffraction detector, grids, plunge freezing for proteins, and trained operators. X-ray source, goniometer, detector and crystal-mounting equipment. Serial delivery hardware or a compatible fixed-target system, plus data-management capacity.
Data completeness Limited by crystal stability and accessible rotation; incomplete angular data can substantially complicate processing. Often benefits from a broad rotation range from one mounted crystal. Completeness comes from combining measurements across crystals and exposures.

Use MicroED when the target forms suitably ordered, sufficiently thin microcrystals and access to cryo-EM expertise is available. Ask about conventional X-ray methods when a larger, stable crystal can be grown. Consider serial X-ray approaches when many small crystals are available and the delivery infrastructure is already established.

Independent reader supportYour contribution helps us test, update, and keep practical guides available for everyone.Support on Ko-Fi

Plan access, materials and sample submission

What the laboratory must provide

  • A defined sample identity and concentration or amount, with any known solvent, buffer, hydrate or polymorph information.
  • Crystals or a crystallization condition that can produce them.
  • Compatibility information for the intended grid and sample state.
  • A realistic amount of material for screening, not merely the theoretical mass contained in one crystal.
  • Agreement on data goals, such as a de novo structure, a ligand pose or confirmation of a known model.

What to ask a facility

  • Does it accept protein microcrystals, small molecules, or both?
  • What crystal dimensions and thicknesses does it prefer?
  • Will staff prepare grids and vitrify protein samples, or must the submitting lab do so?
  • Which detector and data-collection mode are available?
  • How are incomplete data, indexing failures or unusable grids handled?
  • What are the current sample limits, billing rules and scheduling times?

The University of Pennsylvania Electron Microscopy Resource Lab lists a hands-off MicroED service for small-molecule crystals and protein microcrystals. Its service page recommends crystals roughly 50–200 nm in size, accepts one protein sample or four small-molecule samples per submission, and bills instrument time and consumables separately. These are facility-specific terms, not industry-wide requirements; confirm current limits, availability and fees directly before planning a submission.

Special offer. See more information about Outbyte and uninstall instructions. Please review EULA and Privacy policy.

Common failure modes and how to respond

No diffraction from apparently crystalline particles

The particles may be poorly ordered, damaged, too thick or unsuitable for the selected preparation. Re-screen crystallization conditions, compare multiple particles and ask the facility to test thickness and diffraction before collecting a full rotation series.

Best Value
Anker USB C Hub, 5-in-1 USBC to HDMI Splitter with 4K Display | 85W Max PD, 1 Powered USB-C 5Gbps & 2 Powered USB-A 3.0 5Gbps Data Ports, for MacBook Pro, MacBook Air, Dell and More (Black)
  • 5-in-1 Connectivity: Equipped with a 4K HDMI port, a 5 Gbps USB-C data port, two 5 Gbps USB-A ports, and a USB C 100W PD-IN port. Note: The USB C 100W PD-IN port supports only charging and does not support data transfer devices such as headphones or speakers.
  • Powerful Pass-Through Charging: Supports up to 85W pass-through charging so you can power up your laptop while you use the hub. Note: Pass-through charging requires a charger (not included). Note: To achieve full power for iPad, we recommend using a 45W wall charger.
  • Transfer Files in Seconds: Move files to and from your laptop at speeds of up to 5 Gbps via the USB-C and USB-A data ports. Note: The USB C 5Gbps Data port does not support video output.
  • HD Display: Connect to the HDMI port to stream or mirror content to an external monitor in resolutions of up to 4K@30Hz. Note: The USB-C ports do not support video output.
  • What You Get: Anker 332 USB-C Hub (5-in-1), welcome guide, our worry-free 18-month warranty, and friendly customer service.

Protein diffraction degrades after preparation

Dehydration is a primary concern for macromolecules. Review blotting time, humidity control and plunge-freezing conditions. Keep the crystal suspension hydrated until vitrification and avoid treating a protein sample like a dry small-molecule powder.

Ice or support background overwhelms the pattern

Excess liquid can create thick vitreous ice, while an incompatible support film can add background or reduce crystal visibility. Adjust blotting and confirm whether holey-carbon or continuous-carbon grids are appropriate for the sample.

Crystals are trapped against grid bars or clumped together

Grid-bar contact and neighboring crystals make orientation, isolation and rotation difficult. Modify loading density or crystallization handling to obtain better-separated particles. The protein protocol specifically notes that isolated crystals away from grid bars are easier to screen and collect.

Special offer. See more information about Outbyte and uninstall instructions. Please review EULA and Privacy policy.

Data are incomplete or cannot be processed

A few strong diffraction images do not guarantee a solvable dataset. Check the accessible angular range, crystal movement, radiation damage and indexing consistency. Collect from additional crystals when necessary, and plan for the possibility that missing angles will require a new preparation or collection strategy.

A decision checklist for a sub-milligram project

  1. Confirm that the target can form ordered three-dimensional crystals.
  2. Characterize whether the material is a protein, peptide or small molecule, because hydration and temperature requirements differ.
  3. Ask a MicroED facility to review crystal thickness, expected dimensions and sample state before using the full preparation.
  4. Reserve compatible carbon-coated EM grids and, for proteins, a vitrification workflow.
  5. Define how diffraction quality and data completeness will be judged.
  6. Keep a conventional X-ray or serial approach as a contingency when crystal size, quantity or facility access favors it.

MicroED is most powerful when it is treated as a complete workflow rather than a promise based on a tiny mass. Crystal order, suitable thickness, preserved sample state, complete-enough data and experienced support determine whether the final structure is attainable.